POWERPORT CLEARVUE isp — Class II medical device recall Z-2080-2019
Completed recall by Bard Peripheral Vascular Inc, reported 2019-07-31, distributed nationwide. Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack o
| Recall number | Z-2080-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Completed |
| Voluntary or mandated | Voluntary |
| Recalling firm | Bard Peripheral Vascular Inc, Tempe, AZ, United States |
| Recall initiated | 2018-06-25 |
| Classified | 2019-07-25 |
| Reported by FDA | 2019-07-31 |
| Distributed | Nationwide (US) |
| Quantity | 95 units |
Product
POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 6F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 8F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; NAUTILUS DELTA Tip Confirmation System (Includes: Netbook with Pre-loaded Software, Patient Module, and ECG Cable) Product Usage: Is an implantable access device designed to provide repeated access to the vascular system.
Reason for recall
Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA" TCS.