Data Foundry

Medical device recalls 2017

Cholesterol Reagent — Class II medical device recall Z-2081-2017

Terminated recall by Beckman Coulter Inc., reported 2017-05-24, distributed in PR. Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone)

Recall numberZ-2081-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2017-04-03
Classified2017-05-12
Reported by FDA2017-05-24
Terminated2018-05-31
DistributedPR
Countries outside the USAE, AU, BD, BE, BG, CA, CH, CN, CO, CZ, DE, EG, ES, FI, FR, GB, GH, GR, GT, HK, IE, IN, IT, JO, JP, KR, KW, LB, LT, MX, MY, NL, NZ, OM, PA, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SK, TH, TR, TW, VN, ZA
Model numbers467825

Product

Cholesterol Reagent(CHOL 2 x 300), Catalog No. 467825 Intended for quantitative determination of cholesterol concentration in human serum or plasma.

Reason for recall

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.