Data Foundry

Medical device recalls 2015

The FLOW-i Anesthesia System Axillary O2 and Suction Module — Class II medical device recall Z-2083-2015

Terminated recall by Maquet Cardiovascular Us Sales, Llc, reported 2015-07-22, distributed nationwide. Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning f

Recall numberZ-2083-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMaquet Cardiovascular Us Sales, Llc, Wayne, NJ, United States
Recall initiated2015-03-02
Classified2015-07-15
Reported by FDA2015-07-22
Terminated2016-04-29
DistributedNationwide (US)
Countries outside the USAE, AU, BE, BR, CH, CN, CO, CZ, DE, EC, ES, FI, FR, HK, HR, IL, IN, JP, KR, MX, NL, PE, PL, PT, RO, RS, RU, SE, SG, TH, TW, ZA
Quantity104 units
Serial numbers138, 14-24, 199, 205, 26-29, 367, 369-372, 391, 394-395, 416, 466-491, 517, 519, 525, 567-568, 570, 582-583, 606, 613, 619, 627, 629-631, 633, 635-641, 643-665
Model numbers847

Product

The FLOW-i Anesthesia System Axillary O2 and Suction Module, Device Part number: 66 79 847 FLOW-i Anesthesia System: Is indicated for administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.The system is intended for use in a wide range of population: from neonatal to adult patients. The system is intended for use in hospital environments, except for the MRI environment, by healthcare professionals trained in inhalation anesthesia administration. Auxiliary O2 and Suction Module: Is an option available for the FLOW-i Anesthesia System. The module contains two main parts: Flowmeter unit: The intended use of the flowmeter is to provide oxygen for patient therapy. Suction unit: The intended use is to extract body fluids from the stomach and airways.

Reason for recall

Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off switch leading to a decrease or loss of suctioning capability of the modules.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2083-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.