The Drain Bag Set — Class II medical device recall Z-2083-2021
Terminated recall by Fresenius Medical Care Holdings, Inc., reported 2021-07-28, distributed in TX. Exposed to freezing temperatures due to refrigerated truck malfunctions
| Recall number | Z-2083-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Fresenius Medical Care Holdings, Inc., Waltham, MA, United States |
| Recall initiated | 2021-04-28 |
| Classified | 2021-07-16 |
| Reported by FDA | 2021-07-28 |
| Terminated | 2022-08-08 |
| Distributed | TX |
| Quantity | 10 cases |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 00840861102037 |
| Lot numbers | 20NR08163 |
Product
The Drain Bag Set (tubings and flexible bags) is a passive, closed drainage system used as an optional receptacle during an APD treatment. The Drain Bag Set is used to collect effluent in bags rather than letting the effluent flow directly to a drain. The Drain Bag Set connects to the drain line of a cycler set to collect patient effluent by means of gravity and the pumping action of a Peritoneal Dialysis (PD) cycler. The Drain Bag Set connects to the drain line using a standard Luer lock connection. The Drain Bag Set consists of three (3) interconnected 7-liter drain bags, each with a sampling port and an occluding clamp. The tubing lines that connect each drain bag to the drain line have a snap-disconnect tubing segment. This tubing segment allows the user to disconnect individual bags by snapping the tubing apart.
Reason for recall
Exposed to freezing temperatures due to refrigerated truck malfunctions