K-wire — Class II medical device recall Z-2084-2016
Terminated recall by Integra LifeSciences Corp., reported 2016-07-06, distributed nationwide. The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stai
| Recall number | Z-2084-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Integra LifeSciences Corp., Plainsboro, NJ, United States |
| Recall initiated | 2016-05-20 |
| Classified | 2016-06-28 |
| Reported by FDA | 2016-07-06 |
| Terminated | 2016-11-16 |
| Distributed | Nationwide (US) |
| Quantity | 182 units |
| Lot numbers | 237830TE |
Product
K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws
Reason for recall
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.