Data Foundry

Medical device recalls 2016

K-wire — Class II medical device recall Z-2084-2016

Terminated recall by Integra LifeSciences Corp., reported 2016-07-06, distributed nationwide. The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stai

Recall numberZ-2084-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corp., Plainsboro, NJ, United States
Recall initiated2016-05-20
Classified2016-06-28
Reported by FDA2016-07-06
Terminated2016-11-16
DistributedNationwide (US)
Quantity182 units
Lot numbers237830TE

Product

K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws

Reason for recall

The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.