Data Foundry

Medical device recalls 2013

Illico Ti Cannulated Polyaxial Screw 7 — Class II medical device recall Z-2086-2013

Terminated recall by Alphatec Spine, Inc., reported 2013-09-04, distributed nationwide. The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants,

Recall numberZ-2086-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAlphatec Spine, Inc., Carlsbad, CA, United States
Recall initiated2013-06-28
Classified2013-08-27
Reported by FDA2013-09-04
Terminated2014-10-09
DistributedNationwide (US)
Countries outside the USES, FR, JP
Quantity333 screws
Model numbers73875-40

Product

Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.

Reason for recall

The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.