Data Foundry

Medical device recalls 2017

GP321R — Class II medical device recall Z-2088-2017

Terminated recall by Aesculap Implant Systems LLC, reported 2017-05-24, distributed nationwide. Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected

Recall numberZ-2088-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAesculap Implant Systems LLC, Center Valley, PA, United States
Recall initiated2017-03-28
Classified2017-05-12
Reported by FDA2017-05-24
Terminated2018-02-06
DistributedNationwide (US)
Quantity608 units
Reason classesforeign material, specification failure, device malfunction
Model numbersGP321R

Product

GP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING DIAMOND BURR COARSE D4.0; GP324R; ELAN 4 2-RING DIAMOND BURR COARSE D5.0; GP325R; ELAN 4 2-RING DIAMOND BURR COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

Reason for recall

Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.