Data Foundry

Medical device recalls 2012

ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp — Class II medical device recall Z-2089-2012

Ongoing recall by Stryker Howmedica Osteonics Corp., reported 2012-08-01, distributed nationwide. Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Mod

Recall numberZ-2089-2012
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmStryker Howmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2012-04-30
Classified2012-07-26
Reported by FDA2012-08-01
DistributedNationwide (US)
Quantity9,003 units

Product

ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial Estate, Carrigtwohill County Cork, Ireland. The ABG II Modular range of implants is constituted of 8 right stems and 8 left stems. The necks are made of GADS Vitallium, a proprietary Co Cr alloy that Stryker has developed. GADS stands for Gas Atomized Dispersion Strengthened. IT is a Co Cr alloy that is a corrosion resistant and demonstrates qualities of improved fatigue strength (with respect to regular Co Cr alloy). Neck Range - The ABG II modular neck range is composed of 10 reversible neck implants. There are different options in terms of length, version and neck angle. The AG II Modular necks come in two different lengths: 28 mm (short) and 36 mm (long). The necks are offered in two version angles ; 0 and 7. The 0 and 7 necks are available in 125, 130 and 135 neck angles.

Reason for recall

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an