GE426R — Class II medical device recall Z-2089-2017
Terminated recall by Aesculap Implant Systems LLC, reported 2017-05-24, distributed nationwide. Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected
| Recall number | Z-2089-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Aesculap Implant Systems LLC, Center Valley, PA, United States |
| Recall initiated | 2017-03-28 |
| Classified | 2017-05-12 |
| Reported by FDA | 2017-05-24 |
| Terminated | 2018-02-06 |
| Distributed | Nationwide (US) |
| Quantity | 608 |
| Reason classes | foreign material, specification failure, device malfunction |
| Model numbers | GE426R |
Product
GE426R; HI-LINE XS DIAMOND BURR X-CRS.I D4.5MM; GE426SU; HI-LINE XS DISP.DIAM.BURR X-CRS.I D4.5MM; GE526R; HI-LINE XS DIAMOND BURR X-CRS.II D4.5MM; GE526SU; HI-LINE XS DISP.DIAM.BURR X-CRS.II D4.5; GE626R; HI-LINE XS DIAMOND BURR X-CRS.III D4.5MM; GE626SU; HI-LINE XS DISP.DIAM.BURR X-CRS.III D4.5; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Reason for recall
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2089-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.