Data Foundry

Medical device recalls 2023

BD Synapsys Informatics Solution — Class II medical device recall Z-2099-2023

Ongoing recall by Becton Dickinson & Co., reported 2023-07-12, distributed in CA, CO, IA, IL, MI, MN, MO, NY…. There is a risk of mis-association of patient demographic data for users who choose to report results manually

Recall numberZ-2099-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Co., Sparks, MD, United States
Recall initiated2023-06-08
Classified2023-07-05
Reported by FDA2023-07-12
DistributedCA, CO, IA, IL, MI, MN, MO, NY, OH, OK, SC, TN, TX, UT, WI
Countries outside the USAT, AU, BE, CA, CH, DE, DK, ES, FR, GB, IT, JP, LU, NL, NO, NZ, SA, SE, SG, TW, ZA
Quantity102 systems
UPC / GTIN / UDI-DI00382904441500
Model numbers444150

Product

BD Synapsys Informatics Solution, Catalog Number 444150

Reason for recall

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2099-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.