See full list in RES — Class II medical device recall Z-2109-2026
Ongoing recall by Medline Industries, LP, reported 2026-05-13, distributed nationwide. Medline has identified issues related to calibration of the equipment used to sterilize and package the device
| Recall number | Z-2109-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries, LP, Northfield, IL, United States |
| Recall initiated | 2026-01-07 |
| Classified | 2026-05-07 |
| Reported by FDA | 2026-05-13 |
| Distributed | Nationwide (US) |
| Quantity | 15,806 units |
| Reason classes | sterility |
| UPC / GTIN / UDI-DI | 10193489206678, 10193489235890, 10193489261974, 10193489319323, 10193489337112, 10193489394696, 10193489425048, 10193489425536, 10193489475937, 10193489477351, 10193489477436, 10193489477870, 10193489477894, 10193489477955, 10193489478211, 10193489479317, 10193489489316, 10193489497489, 10193489601794, 10193489648867, 10193489685190, 10193489704921, 10193489779257, 10193489795516, 10193489814163 and 177 more |
| Lot numbers | 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 117, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137, 138, 139 and 460 more |
| Model numbers | CDS981504G, CDS982407P, CDS982893I, CDS983300J, CDS984243G, CDS984543D, CDS984543F, CDS984543G, CDS984543I, CDS984860F, CDS985423G, DYKM1657A, DYKMBNDL100, DYKMBNDL109, DYKMBNDL116D, DYKMBNDL116F, DYKMBNDL119, DYKMBNDL119A, DYKMBNDL168, DYKMBNDL168A, DYKMBNDL184D, DYKMBNDL91 |
Product
See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;
Reason for recall
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Machine access