FlexLab Automation System used with the following Modules: High… — Class II medical device recall Z-2114-2020
Terminated recall by Inpeco S.A., reported 2020-06-03, distributed in IL, NY. The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the
| Recall number | Z-2114-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Inpeco S.A., Lugano, N/A, Switzerland |
| Recall initiated | 2020-04-07 |
| Classified | 2020-05-26 |
| Reported by FDA | 2020-06-03 |
| Terminated | 2024-06-13 |
| Distributed | IL, NY |
| Quantity | 10 systems |
| Serial numbers | FLX.0133, FLX.0181, FLX.0191, FLX.0192, FLX.0197, FLX.0198, FLX.0200, FLX.0201, FLX.0202, FLX.027 |
Product
FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Reason for recall
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2114-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.