Data Foundry

Medical device recalls 2020

Accelerator a3600 Automation System used with the following Modules:… — Class II medical device recall Z-2115-2020

Terminated recall by Inpeco S.A., reported 2020-06-03, distributed in IL, NY. The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the

Recall numberZ-2115-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInpeco S.A., Lugano, N/A, Switzerland
Recall initiated2020-04-07
Classified2020-05-26
Reported by FDA2020-06-03
Terminated2024-06-13
DistributedIL, NY
Quantity2 systems
Serial numbersACP.050, ACP.146

Product

Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason for recall

The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.