Data Foundry

Medical device recalls 2013

GE Healthcare — Class II medical device recall Z-2117-2013

Terminated recall by GE Healthcare, LLC, reported 2013-09-04, distributed nationwide. GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Re

Recall numberZ-2117-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2013-04-03
Classified2013-08-28
Reported by FDA2013-09-04
Terminated2016-07-25
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GR, HK, HN, HU, ID, IL, IN, IQ, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MX, MY, NG, NL, NO, NP, NZ, PA, PE, PH, PL, PT, PY, QA, RO, RU, SA, SE, SG, SI, SV, TH, TR, TT, TW, UY, VE, ZA
Serial numbersRT406294095GR, RT407035566GR, RT410149277GA, RT410149280GA, RT410149288GA, RT410149301GAMKE, RT410149303GA, RT410189351GA, RT410229385GA, RT410259423GA, RT410259424GA, RT410259425GA, RT410289453GA, RT410289457GA, RT410289460GA, RT410289481GA, RT410289483GA, RT410289486GA, RT410309514GA, RT410329525GA, RT410329532GA, RT410329542GA, RT410329546GA, RT410329547GA, RT410329548GA and 475 more

Product

GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another.

Reason for recall

GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2117-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.