Medtronic Nexdrive Micropositioning Drive — Class II medical device recall Z-2118-2014
Terminated recall by Medtronic Neuromodulation, reported 2014-08-06, distributed in CA, CO, FL, IL, KS, KY, LA, MN…. Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in th
| Recall number | Z-2118-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2014-06-11 |
| Classified | 2014-07-25 |
| Reported by FDA | 2014-08-06 |
| Terminated | 2015-01-22 |
| Distributed | CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, UT, VA |
| Countries outside the US | CA, ES, GB |
| Lot numbers | 082601414, 082602014, 082605614, 082616214, 082616314, 082618213, 082618314, 082618414, 082621413, 082623113, 082626913, 082629613, 082629713, 082630913, 082634313 |
| Model numbers | ML-1000, ML-2000 |
Product
Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.
Reason for recall
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.