10-1335M — Class I medical device recall Z-2118-2017
Terminated recall by EBI Patient Care, Inc., reported 2017-05-31, distributed nationwide. Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spi
| Recall number | Z-2118-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | EBI Patient Care, Inc., Guaynabo, PR, United States |
| Recall initiated | 2017-04-20 |
| Classified | 2017-05-25 |
| Reported by FDA | 2017-05-31 |
| Terminated | 2018-10-17 |
| Distributed | Nationwide (US) |
| Quantity | 33 units |
| Serial numbers | 224595, 224598, 224607, 224608, 224610, 224613, 224615, 224621, 224622, 224623, 224624, 224625, 224626, 224644, 224649, 224651, 224655, 224656, 224658, 224659, 224666, 224667 |
Product
10-1335M - SpF-XL IIB 2/DM Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels.
Reason for recall
Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher than allowed cytotoxicity levels found during routine process monitoring with the product.