Data Foundry

Medical device recalls 2025

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry… — Class II medical device recall Z-2121-2025

Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2025-07-23, distributed nationwide. Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of chol

Recall numberZ-2121-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Inc., Rochester, NY, United States
Recall initiated2025-06-10
Classified2025-07-17
Reported by FDA2025-07-23
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
Reason classeslabeling, specification failure
UPC / GTIN / UDI-DI10758750004577
Model numbers8231474
Expiration dates2025-08-14, 2025-11-20, 2026-03-25, 2026-06-03, 2027-02-03

Product

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier II Model/Catalog Number: 8231474 Software Version: N/A Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component

Reason for recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2121-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.