Data Foundry

Medical device recalls 2025

VITROS Chemistry Products CHE Slides — Class II medical device recall Z-2123-2025

Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2025-07-23, distributed nationwide. Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of chol

Recall numberZ-2123-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Inc., Rochester, NY, United States
Recall initiated2025-06-10
Classified2025-07-17
Reported by FDA2025-07-23
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
Quantity414,000
Reason classeslabeling, specification failure
UPC / GTIN / UDI-DI10758750030293, 10758750030309
Model numbers10758750030293, 10758750030309, 4074, 4138, 684

Product

VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Reason for recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).