Evans Plate — Class II medical device recall Z-2127-2018
Terminated recall by Orthofix, Inc, reported 2018-06-13, distributed in CA, DC, FL, GA, IL, IN, LA, MD…. Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating Syste
| Recall number | Z-2127-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Orthofix, Inc, Lewisville, TX, United States |
| Recall initiated | 2018-04-06 |
| Classified | 2018-06-07 |
| Reported by FDA | 2018-06-13 |
| Terminated | 2021-06-30 |
| Distributed | CA, DC, FL, GA, IL, IN, LA, MD, MI, MN, NC, NJ, NV, OH, OK, PA, SC, TX, VA |
| Quantity | 52 |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 18056099647724, 18056099647731, 18056099647748, 18056099647755 |
| Lot numbers | B1212604 |
| Model numbers | 6MM |
| Expiration dates | 2022-10-15 |
Product
Evans Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 6MM Spacer (002-H-00006); b) 8MM Spacer (022-H-00008); c) 10MM Spacer (002-H-00010); d) 12MM Spacer (002-H-00012)
Reason for recall
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.