VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use… — Class II medical device recall Z-2129-2023
Ongoing recall by Ortho-Clinical Diagnostics, reported 2023-07-19, distributed nationwide. Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of
| Recall number | Z-2129-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2023-05-08 |
| Classified | 2023-07-11 |
| Reported by FDA | 2023-07-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG |
| Quantity | 2,400 units |
| UPC / GTIN / UDI-DI | 10758750006687 |
| Lot numbers | 2022 |
| Expiration dates | 2023-09-06 |
Product
VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM. Product Code: 6801704
Reason for recall
Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2129-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.