Data Foundry

Medical device recalls 2020

Philips Sterilizable Defibrillator Internal Paddles — Class II medical device recall Z-2134-2020

Terminated recall by Philips North America, LLC, reported 2020-06-03, distributed nationwide. Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instruc

Recall numberZ-2134-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, LLC, Andover, MA, United States
Recall initiated2020-04-20
Classified2020-05-27
Reported by FDA2020-06-03
Terminated2021-04-02
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DO, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HT, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, LB, LK, LV, MA, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SK, SV, TH, TN, TR, TW, UA, VE, VN, ZA
Quantity21,683 internal

Product

Philips Sterilizable Defibrillator Internal Paddles - Internal paddles are an accessory to the compatible monitor/defibrillator Switchless: M1741A, M1742A, M1743A, M1744A Switched: M4741A, M4742A, M4743A, M4744A - Product Usage: An electrical conductor used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.

Reason for recall

Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather than specifying a maximum number of sterilization cycles to determine when the paddles should be discarded. A damaged Internal Paddle may not be able to deliver therapy.