Data Foundry

Medical device recalls 2024

CORI Starter Kit — Class II medical device recall Z-2135-2024

Ongoing recall by Blue Belt Technologies, Inc, reported 2024-06-19, distributed nationwide. CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) whi

Recall numberZ-2135-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBlue Belt Technologies, Inc, Pittsburgh, PA, United States
Recall initiated2024-05-07
Classified2024-06-11
Reported by FDA2024-06-19
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, BR, CA, CL, CN, CO, CY, DE, ES, FR, GB, GR, HK, IE, IL, IN, IT, JP, KR, KW, MY, NL, NO, NZ, PL, PT, SA, SE, SG, TH, TR, TW, ZA
Quantity916 units
Reason classessoftware
UPC / GTIN / UDI-DI00885556755068, 00885556757420, 00885556796849
Serial numbersSN000001, SN000002, SN000003, SN000004, SN000005, SN000006, SN000007, SN000008, SN000009, SN000010, SN000011, SN000012, SN000013, SN000014, SN000015, SN000016, SN000017, SN000018, SN000019, SN000020, SN000021, SN000022, SN000023, SN000024, SN000025 and 475 more

Product

CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures Product Numbers: (1) ROB10000 CORI Starter Kit (2) ROB10024 Real Intelligence CORI (3) ROB10197 CORI STARTER KIT - UNIVERSAL Note: the three product numbers cover the three different configurations in which the CORI Console is sold

Reason for recall

CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to flicker on screen of the CORI Surgical System during use, and may cause surgical delay