Data Foundry

Medical device recalls 2025

HSG Procedure Tray — Class II medical device recall Z-2135-2025

Ongoing recall by CooperSurgical, Inc., reported 2025-07-30, distributed nationwide. The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the pe

Recall numberZ-2135-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2025-06-11
Classified2025-07-18
Reported by FDA2025-07-30
DistributedNationwide (US)
Quantity171 units
Reason classeslabeling, sterility
UPC / GTIN / UDI-DI60888937003262
Lot numbers339327, 339331, 339333, 339334, 343007
Model numbers6050T

Product

HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterography procedures.

Reason for recall

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2135-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.