Data Foundry

Medical device recalls 2024

CODMAN EVD BACTISEAL 1 — Class II medical device recall Z-2137-2024

Ongoing recall by Integra LifeSciences Corp., reported 2024-06-19, distributed nationwide. Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise t

Recall numberZ-2137-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corp., Princeton, NJ, United States
Recall initiated2024-05-07
Classified2024-06-12
Reported by FDA2024-06-19
DistributedNationwide (US)
Countries outside the USCA, CN, HR, HU, RU, SG, SI, TR, TW, ZA
Quantity106 units
Reason classessterility, packaging
UPC / GTIN / UDI-DI10381780518440
Lot numbers7314912, 7320010, 7322933

Product

CODMAN EVD BACTISEAL 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume. Product Number: 821749

Reason for recall

Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2137-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.