ABL90 FLEX — Class II medical device recall Z-2139-2020

Terminated recall by Radiometer Medical ApS, reported 2020-06-03, distributed nationwide. The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the opt

Recall numberZ-2139-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRadiometer Medical ApS, Bronshoj, N/A, Denmark
Recall initiated2019-11-21
Classified2020-05-28
Reported by FDA2020-06-03
Terminated2024-11-08
DistributedNationwide (US)
Quantity1,233 units
Model numbers393-090, R0659, R0977

Product

ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for recall

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.