Data Foundry

Medical device recalls 2016

The LifeCare Pump Model 4200 PCA Infuser makes possible the practical… — Class II medical device recall Z-2142-2016

Terminated recall by Hospira Inc., reported 2016-07-20, distributed nationwide. Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious

Recall numberZ-2142-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHospira Inc., Lake Forest, IL, United States
Recall initiated2013-03-08
Classified2016-07-08
Reported by FDA2016-07-20
Terminated2016-11-07
DistributedNationwide (US)
Quantity6,268 infusion
Reason classesdevice malfunction
Serial numbers100, 1000, 10000, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 1001, 10010, 10011, 10012, 10013, 10014, 10015, 10016, 10017, 10018, 10019, 1002, 10020 and 475 more
Model numbers4200PCA

Product

The LifeCare Pump Model 4200 PCA Infuser makes possible the practical implementation of patient-controlled analgesia. The health care professional sets the prescribed dose, rates, and limits for the drug to be infused. The infuser can be set to operate in one of three modes: PCA Mode, Continuous Mode, and PCA + Continuous Mode. The LifeCare Pump Model 4200 PCA Infuser is designed with a number of mechanical, electrical, and software safeguards that prevent patient injury or drug abuse.

Reason for recall

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2142-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.