Data Foundry

Medical device recalls 2015

ATTUNE CONV FB CR ARTICULATION SURFACE SZ1 INTENDED USE: The Tibial… — Class II medical device recall Z-2145-2015

Terminated recall by DePuy Orthopaedics, Inc., reported 2015-07-29, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Arti

Recall numberZ-2145-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDePuy Orthopaedics, Inc., Warsaw, IN, United States
Recall initiated2015-06-12
Classified2015-07-20
Reported by FDA2015-07-29
Terminated2015-11-12
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE, AT, AU, BE, CA, CH, CL, CO, DE, ES, FR, GB, HK, ID, IE, IL, IN, IT, JP, KR, MX, MY, NL, NZ, PE, SE, SG, SK, TH, ZA
Quantity3,474 units
UPC / GTIN / UDI-DI10603295135388, 10603295135395

Product

ATTUNE CONV FB CR ARTICULATION SURFACE SZ1 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.

Reason for recall

Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.