Data Foundry

Medical device recalls 2016

Optical Surface Monitoring System{OSMS — Class II medical device recall Z-2149-2016

Terminated recall by Varian Medical Systems, Inc., reported 2016-07-20, distributed nationwide. Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, spe

Recall numberZ-2149-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVarian Medical Systems, Inc., Palo Alto, CA, United States
Recall initiated2016-06-23
Classified2016-07-11
Reported by FDA2016-07-20
Terminated2018-08-24
DistributedNationwide (US)
Countries outside the USAT, BG, CH, DE, DK, GB, IN, IT, MA, MY, NO, PT
Serial numbersHOM0003, HOM0004, HOM0005, HOM0006, HOM0008, HOM0009, HOM0010, HOM0011, HOM0012, HOM0015, HOM0016, HOM0017, HOM0018, HOM0019, HOM0020, HOM0021, HOM0022, HOM0023, HOM0024, HOM0025, HOM0026, HOM0027, HOM0028, HOM0029, HOM0030 and 177 more

Product

Optical Surface Monitoring System{OSMS, Varian Private Label} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to Withhold the beam automatically during radiation delivery.

Reason for recall

Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with Optical Surface Monitoring System [OSMS].

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2149-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.