Data Foundry

Medical device recalls 2023

Cardiosave Hybrid Intra-Aortic Balloon Pump — Class I medical device recall Z-2149-2023

Ongoing recall by Datascope Corp., reported 2023-07-26, distributed nationwide. IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batterie

Recall numberZ-2149-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDatascope Corp., Mahwah, NJ, United States
Recall initiated2023-06-05
Classified2023-07-19
Reported by FDA2023-07-26
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BB, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GT, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MX, MY, NG, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, VN, ZA
UPC / GTIN / UDI-DI10607567108391, 10607567108414, 10607567108421, 10607567108438, 10607567109008, 10607567109053, 10607567109107, 10607567111117, 10607567111940, 10607567113432
Model numbers0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 10607567108391, 10607567108414, 10607567108421, 10607567108438, 10607567109008, 10607567109053, 10607567109107, 10607567111117, 10607567111940, 10607567113432

Product

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution

Reason for recall

IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.