PopLok Punch labeled in part: "PKL-35M***3 — Class II medical device recall Z-2151-2012
Terminated recall by Linvatec Corp. dba ConMed Linvatec, reported 2012-08-15. ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm wh
| Recall number | Z-2151-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Linvatec Corp. dba ConMed Linvatec, Largo, FL, United States |
| Recall initiated | 2012-07-06 |
| Classified | 2012-08-06 |
| Reported by FDA | 2012-08-15 |
| Terminated | 2013-11-21 |
| Countries outside the US | KR |
| Quantity | 2 |
| Lot numbers | 318455 |
Product
PopLok Punch labeled in part: "PKL-35M***3.5mm***QTY 1***PopLok Punch***ConMed Linvatec Biomaterials, Ltd. Hermiankatu 6-8L 33720 Tampere, FINLAND***Rx Only***ConMed Linvatec 11311 Concept Blvd Largo, FL 33773-4908 USA***Made in USA***" Intended Use: Bone Punch.
Reason for recall
ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm when it should have been 3.5mm.