SENSATION PLUS¿ 8Fr — Class II medical device recall Z-2164-2017
Terminated recall by Datascope Corporation, reported 2017-05-31, distributed nationwide. Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50c
| Recall number | Z-2164-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datascope Corporation, Fairfield, NJ, United States |
| Recall initiated | 2017-04-12 |
| Classified | 2017-05-25 |
| Reported by FDA | 2017-05-31 |
| Terminated | 2020-01-28 |
| Distributed | Nationwide (US) |
| Quantity | 996 units |
| Reason classes | packaging |
| Lot numbers | 3000043374, 3000045513 |
| Model numbers | 0684-00-0576-01U |
Product
SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.
Reason for recall
Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2164-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.