SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use… — Class II medical device recall Z-2166-2017
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2017-06-07, distributed nationwide. Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stabili
| Recall number | Z-2166-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States |
| Recall initiated | 2017-01-31 |
| Classified | 2017-05-26 |
| Reported by FDA | 2017-06-07 |
| Terminated | 2018-05-24 |
| Distributed | Nationwide (US) |
| Quantity | 9,100 kits |
| Reason classes | specification failure |
Product
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Reason for recall
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.