The VITEK¿ 2 Anaerobic and Corynebacteria identification card — Class II medical device recall Z-2166-2018
Terminated recall by BioMerieux SA, reported 2018-06-20. A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. Afte
| Recall number | Z-2166-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | BioMerieux SA, Marcy L'Etoile, N/A, France |
| Recall initiated | 2018-03-23 |
| Classified | 2018-06-12 |
| Reported by FDA | 2018-06-20 |
| Terminated | 2020-06-23 |
| Reason classes | temperature control |
| Lot numbers | 2440318203 |
| Model numbers | 21347 |
Product
The VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identification of most clinically significant anaerobic organisms and Corynebatcerium species. The VITEK¿ 2 ANC identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Reason for recall
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2166-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.