Data Foundry

Medical device recalls 2026

Philips Azurion — Class II medical device recall Z-2170-2026

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2026-05-20, distributed nationwide. Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor c

Recall numberZ-2170-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands
Recall initiated2026-03-31
Classified2026-05-12
Reported by FDA2026-05-20
DistributedNationwide (US)
Countries outside the USCA
Quantity1,718 units
Reason classesdevice malfunction
UPC / GTIN / UDI-DI00884838085251, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272, 00884838110564, 00884838110571, 00884838116726, 00884838116733, 00884838116740, 00884838116757, 00884838116764, 00884838116771, 00884838116788, 00884838116801

Product

Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).

Reason for recall

Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.