Data Foundry

Medical device recalls 2024

The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System — Class II medical device recall Z-2175-2024

Ongoing recall by Maquet Cardiovascular, LLC, reported 2024-07-03, distributed nationwide. Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device

Recall numberZ-2175-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMaquet Cardiovascular, LLC, Wayne, NJ, United States
Recall initiated2024-05-17
Classified2024-06-24
Reported by FDA2024-07-03
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CN, GB, IN, KR, NZ, TH, TR, TW
Quantity430,037 units
UPC / GTIN / UDI-DI00607567700406, 00607567700451
Model numbersVH-4000, VH-4001

Product

The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Reason for recall

Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2175-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.