Data Foundry

Medical device recalls 2020

Power supply provided with Afinion 2 — Class II medical device recall Z-2192-2020

Ongoing recall by Abbott Diagnostics Technologies AS, reported 2020-06-10, distributed in GA, NJ. The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the ana

Recall numberZ-2192-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbbott Diagnostics Technologies AS, Oslo, N/A, Norway
Recall initiated2020-03-23
Classified2020-05-29
Reported by FDA2020-06-10
DistributedGA, NJ
Countries outside the USAR, AU, BD, BR, CA, CH, CL, CN, CO, DE, DK, ES, FI, FR, GB, GT, HK, IN, IS, IT, JP, KR, LK, MX, MY, NL, NO, OM, PE, PH, PK, PL, SA, SE, SG, TH, TT, TW, ZA
Quantity42,237 instruments
Model numbers1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970 and 4 more

Product

Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.

Reason for recall

The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2192-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.