Data Foundry

Medical device recalls 2019

Philips EPIQ and Affiniti Ultrasound Systems with software version 4 — Class II medical device recall Z-2199-2019

Terminated recall by Philips Ultrasound Inc, reported 2019-08-14, distributed nationwide. There is a potential that the image of one patient could get unexpectedly moved into another patient's folder

Recall numberZ-2199-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Ultrasound Inc, Bothell, WA, United States
Recall initiated2019-04-29
Classified2019-08-06
Reported by FDA2019-08-14
Terminated2022-08-04
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BH, BO, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HN, HR, HU, ID, IL, IN, IR, IT, JP, LK, LU, LV, MA, MX, MY, NI, NL, NO, NZ, OM, PE, PH, PL, PT, QA, RO, RS, RU, SA, SE, SG, TH, TR, TW, UY, VN, ZA
Quantity3,146 devices

Product

Philips EPIQ and Affiniti Ultrasound Systems with software version 4.0: Models EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIZ CVx, Affiniti 30, Affiniti 50, and Affiniti 70. Product Usage: The intended use of the EPIQ, EPIQ 5, EPIQ 7, Affiniti 30, Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Or Jan (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal. When integrated with Philips EchoNavigator, the systems can assist the interventionist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.

Reason for recall

There is a potential that the image of one patient could get unexpectedly moved into another patient's folder when the user is using the Edit functionality on the Patient Data Entry (PDE) screen of the device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2199-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.