T8 REMOVAL HEAD — Class II medical device recall Z-2200-2013

Terminated recall by Medtronic Sofamor Danek USA Inc, reported 2013-09-25, distributed nationwide. One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head

Recall numberZ-2200-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Sofamor Danek USA Inc, Memphis, TN, United States
Recall initiated2013-08-12
Classified2013-09-14
Reported by FDA2013-09-25
Terminated2013-12-27
DistributedNationwide (US)
Quantity9 units
Model numbers8910641

Product

T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.

Reason for recall

One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.