Data Foundry

Medical device recalls 2019

LSB Style Anchor Bolt — Class II medical device recall Z-2204-2019

Terminated recall by Ad-Tech Medical Instrument Corporation, reported 2019-08-14, distributed nationwide. On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 20

Recall numberZ-2204-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAd-Tech Medical Instrument Corporation, Oak Creek, WI, United States
Recall initiated2019-06-18
Classified2019-08-06
Reported by FDA2019-08-14
Terminated2021-02-16
DistributedNationwide (US)
Countries outside the USAU
Quantity201 devices
Reason classeslabeling, sterility
Model numbers13MM

Product

LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 13mm non-anodized anchor bolt body with anchor bolt cap containing a green gasket and silicone cap, 26mm non-anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap - Product Usage: The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.

Reason for recall

On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2204-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.