HX2" Temperature Management Systems — Class II medical device recall Z-2207-2016
Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2016-07-27, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due t
| Recall number | Z-2207-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Terumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States |
| Recall initiated | 2016-06-08 |
| Classified | 2016-07-18 |
| Reported by FDA | 2016-07-27 |
| Terminated | 2017-08-04 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY |
| Countries outside the US | AE, AR, AU, BE, BR, CA, CH, CL, CN, CO, CR, DE, DO, EC, ES, FI, FR, GB, GR, HK, ID, IE, IL, IN, IR, IT, JM, JO, JP, KR, LB, MT, MX, MY, NL, NO, NZ, PH, PT, RU, SA, SE, SG, SV, TH, TR, TW, UY, VE, VN |
| Quantity | 6,412 units |
| Model numbers | 809810 |
Product
HX2" Temperature Management Systems; The Terumo¿ HX2" Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right).The system has the capacity to circulate water at a rate of up to 6.5 gal.min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Product Usage: HX2 Temperature Management Systems is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.