Data Foundry

Medical device recalls 2025

Ingenia Ambition S Model Numbers — Class II medical device recall Z-2208-2025

Ongoing recall by Philips North America, reported 2025-08-06, distributed nationwide. The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source

Recall numberZ-2208-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, Cambridge, MA, United States
Recall initiated2025-06-30
Classified2025-07-30
Reported by FDA2025-08-06
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BG, BH, BO, BR, BZ, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LU, LV, MA, MT, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TT, TW, UA, VE, VN, ZA
Quantity611 units
UPC / GTIN / UDI-DI00884838090057, 00884838098343, 00884838108639
Serial numbers27500, 27502, 27504, 27505, 27506, 27507, 27508, 27509, 27511, 27512, 27514, 27515, 27516, 27517, 27518, 27519, 27520, 27521, 27522, 27523, 27524, 27525, 27526, 27527, 27528 and 475 more
Model numbers00884838090057, 00884838098343, 781359, 782108, 782133, 782139

Product

Ingenia Ambition S Model Numbers (REF): (1) 781359, (2) 782108, (3) 782133 (China ONLY), (4) 782139;

Reason for recall

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2208-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.