Data Foundry

Medical device recalls 2012

COBAS AmpliPrep Sample Tube Input Barcode Clips — Class II medical device recall Z-2209-2012

Terminated recall by Roche Molecular Systems, Inc., reported 2012-08-22, distributed nationwide. Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause th

Recall numberZ-2209-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Molecular Systems, Inc., Branchburg, NJ, United States
Recall initiated2012-02-01
Classified2012-08-15
Reported by FDA2012-08-22
Terminated2013-07-16
DistributedNationwide (US)
Quantity2,132 boxes
Lot numbers11049503, 11053503, 11054503, 11525503, 11526503
Model numbers03137040001

Product

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Reason for recall

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce