Data Foundry

Medical device recalls 2016

Sarns TCM II — Class II medical device recall Z-2209-2016

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2016-07-27, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due t

Recall numberZ-2209-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States
Recall initiated2016-06-08
Classified2016-07-18
Reported by FDA2016-07-27
Terminated2017-08-04
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAE, AR, AU, BE, BR, CA, CH, CL, CN, CO, CR, DE, DO, EC, ES, FI, FR, GB, GR, HK, ID, IE, IL, IN, IR, IT, JM, JO, JP, KR, LB, MT, MX, MY, NL, NO, NZ, PH, PT, RU, SA, SE, SG, SV, TH, TR, TW, UY, VE, VN
Quantity6,412 units

Product

Sarns TCM II; The Sarns Temperature Control and Monitor unit (TCM) II (system) is a source of temperature controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. Product Usage: The Sarns TCM II is indicated for controlling and monitoring patient temperature.

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2209-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.