Data Foundry

Medical device recalls 2026

Automated Impella Controller — Class I medical device recall Z-2211-2026

Ongoing recall by Abiomed, Inc., reported 2026-06-03, distributed nationwide. A retrospective review identified that there are AIC units requiring specific hardware updates to address pote

Recall numberZ-2211-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbiomed, Inc., Danvers, MA, United States
Recall initiated2026-04-20
Classified2026-05-22
Reported by FDA2026-06-03
DistributedNationwide (US)
Countries outside the USAE, BR, CA, CO, GB, HK, IT, KW, SA
Quantity91 units
UPC / GTIN / UDI-DI00813502010022
Serial numbersIC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1301, IC1390, IC1544, IC1545, IC1572, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278 and 66 more

Product

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

Reason for recall

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2211-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.