Zeppessis' reprocessed single-use manifolds — Class II medical device recall Z-2212-2013
Terminated recall by Zeppessis Reprocessing, LLC, reported 2013-09-25, distributed in CA, ID, OR, WA. Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management
| Recall number | Z-2212-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zeppessis Reprocessing, LLC, Garden City, ID, United States |
| Recall initiated | 2013-08-16 |
| Classified | 2013-09-17 |
| Reported by FDA | 2013-09-25 |
| Terminated | 2013-11-08 |
| Distributed | CA, ID, OR, WA |
| Model numbers | 700-015-000, 700-020-000, 702-020-000, 702-025-000 |
Product
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 1 (part # 700-020-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***single-port manifold for the Stryker Neptune 2 (part # 702-025-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 2 (part # 702-020-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump.
Reason for recall
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol for reprocessing the used manifolds.