Data Foundry

Medical device recalls 2012

00-4309-028-01 — Class II medical device recall Z-2213-2012

Terminated recall by Zimmer, Inc., reported 2012-08-22, distributed nationwide. Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of align

Recall numberZ-2213-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2012-07-18
Classified2012-08-16
Reported by FDA2012-08-22
Terminated2014-04-17
DistributedNationwide (US)
Countries outside the USBE, CH, DE, DK, ES, FI, FR, GB, IT, NL, SE
Quantity252
Reason classesdevice malfunction
Lot numbers60529860, 60574377, 60589079, 60605856, 60616574, 60684774, 60739623, 60842967, 60896979, 60936815, 61019785, 61051017
Model numbers00-4309-028-01

Product

00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 40mm 65¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."

Reason for recall

Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2213-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.