YAMATO PLUS IAB — Class II medical device recall Z-2216-2023
Ongoing recall by Datascope Corp., reported 2023-08-02, distributed nationwide. During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introdu
| Recall number | Z-2216-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datascope Corp., Fairfield, NJ, United States |
| Recall initiated | 2023-06-23 |
| Classified | 2023-07-21 |
| Reported by FDA | 2023-08-02 |
| Distributed | Nationwide (US) |
| Quantity | 321,609 kits |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 10607567107585, 10607567107592, 10607567107608, 10607567107615, 10607567107622, 10607567107639 |
Product
YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description Part Number YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0559-01 YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0559-02 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0560-01 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0560-02 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0561-01 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0561-02 PACKAGED INSERTION KIT - YAMATO PLUS-R 7.5 Fr. 30/35/40 cc IABs D886-00-0019-25
Reason for recall
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.