TRANS-RAY IAB — Class II medical device recall Z-2218-2023
Ongoing recall by Datascope Corp., reported 2023-08-02, distributed nationwide. During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introdu
| Recall number | Z-2218-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datascope Corp., Fairfield, NJ, United States |
| Recall initiated | 2023-06-23 |
| Classified | 2023-07-21 |
| Reported by FDA | 2023-08-02 |
| Distributed | Nationwide (US) |
| Quantity | 321,609 kits |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 10607567107684, 10607567111711, 10607567111728, 10607567111810 |
| Model numbers | 7FR |
Product
TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26
Reason for recall
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.