Data Foundry

Medical device recalls 2017

Exact Sciences Sample Mixer 2 — Class II medical device recall Z-2219-2017

Terminated recall by Exact Sciences Corporation, reported 2017-06-07, distributed in WI. Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damag

Recall numberZ-2219-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmExact Sciences Corporation, Madison, WI, United States
Recall initiated2017-04-28
Classified2017-06-01
Reported by FDA2017-06-07
Terminated2017-08-11
DistributedWI
Quantity15
Serial numbersMIX-0051, MIX-0052, MIX-0053, MIX-0054, MIX-0055, MIX-0056, MIX-0057, MIX-0058, MIX-0059, MIX-0060, MIX-0061, MIX-0063, MIX-0064, MIX-0065, MIX-0066, MX00002014, MX00002015, MX00002016, MX00002027, MX00002028, MX00002032, MX00002033, MX00002036, MX00002039, MX00002040 and 5 more

Product

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Reason for recall

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2219-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.