Data Foundry

Medical device recalls 2025

Volcano Visions Digital IVUS Catheter: PV — Class II medical device recall Z-2221-2025

Ongoing recall by Volcano Corp, reported 2025-08-06, distributed nationwide. Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter u

Recall numberZ-2221-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmVolcano Corp, San Diego, CA, United States
Recall initiated2025-06-13
Classified2025-07-30
Reported by FDA2025-08-06
DistributedNationwide (US)
Quantity233,817
UPC / GTIN / UDI-DI00184360000037, 00845225002848
Model numbers014R, 85910P, 86700, 86700J

Product

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Reason for recall

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2221-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.